Opentrons Flex Gripper - automated labware movement between deck locations, modules, waste chute, and off-deck storage with precise positioning and offset control for hands-free plate transfers

The file

Show its rules and text

We split this skill into the rules it gives an agent

29 rules. About 0 are common, 24 are distinctive, and 5 are very specific to this project.

  • ≈0

    Common · estimate

    Many files like this have a rule of this kind.

  • ≈24

    Distinctive · estimate

    This author's own choices. Worth a look.

  • 5

    Very specific

    Tied to their project's names, files or tools. Probably skip.

  • Common
  • Distinctive
  • Very specific
  1. Distinctive:

    Use the Gripper skill when:

    line 19
  2. Very specific:

    Robot type: Must specify "robotType": "Flex"

    line 48
  3. Distinctive:

    Ensure proper gripper engagement

    line 154
  4. Distinctive:

    Ensure proper seating on module or adapter

    line 157
  5. Distinctive:

    Avoid collision with deck features

    line 159
  6. Very specific:

    Solution: Use use_gripper=False to prompt manual movement, or use adapters.

    line 439
  7. Distinctive:

    Test with simulation first: - Verify gripper movements before running on hardware

    line 445
  8. Distinctive:

    Add protocol comments: - Document why labware is being moved

    line 446
  9. Distinctive:

    Check labware compatibility: - Ensure labware is grippable before deploying

    line 449
  10. Distinctive:

    Use offsets judiciously: - Only when necessary for proper positioning

    line 450
  11. Distinctive:

    Dispose properly: - Use waste chute for used consumables to maintain workspace

    line 452
  12. Distinctive:

    Verify labware has gripper-compatible features

    line 517
  13. Distinctive:

    Check labware is properly seated on deck/module

    line 518
  14. Very specific:

    Use use_gripper=False for non-compatible labware

    line 520
  15. Distinctive:

    Ensure module lids are open (Thermocycler, Absorbance Reader)

    line 523
  16. Distinctive:

    Verify Heater-Shaker latch is open

    line 524
  17. Distinctive:

    Check deck layout for obstructions

    line 525
  18. Distinctive:

    Check module/adapter compatibility with labware

    line 530
  19. Distinctive:

    Verify destination surface is level and clear

    line 531
  20. Very specific:

    Check that use_gripper=True is specified

    line 534
  21. Distinctive:

    Verify gripper is enabled for protocol

    line 535
  22. Distinctive:

    Ensure labware is grippable

    line 536
  23. Very specific:

    Import OFF_DECK constant: from opentrons.protocol_api import OFF_DECK

    line 539
  24. Distinctive:

    Ensure sufficient off-deck storage slots

    line 541
  25. Distinctive:

    Keep high-traffic locations accessible: - Place frequently moved labware in central deck positions

    line 606
  26. Distinctive:

    Reserve column 4 for Absorbance Reader: - If using reader, entire column 4 is staging area

    line 608
  27. Distinctive:

    Plan waste chute access: - Ensure gripper can reach waste chute without obstacles

    line 610
  28. Distinctive:

    Important: When Absorbance Reader is loaded, column 4 cannot be used for labware.

    line 628
  29. Distinctive:

    Robot type: Must be Opentrons Flex

    line 634

≈ “Common” is an estimate: at least 1 in 5 of the 350,495 skill files we read have a rule of the same kind, not the same words. “Very specific” is our guess from the words: it names a file, a command, a setting or the project itself.

Keep what fits and drop the rest. Your version, ready for your agents.
Files
SKILL.md

Opentrons Flex Gripper

Overview

The Opentrons Flex Gripper enables fully automated labware movement without manual intervention. It moves plates, reservoirs, and tip racks between deck slots, hardware modules, trash/waste chute, and off-deck storage with precise positioning and optional offset adjustments.

Core value: Eliminate manual plate transfers during protocols. Automatically move labware between modules, enabling complex multi-step workflows without pausing for user intervention.

Platform: Opentrons Flex only (not available on OT-2)

When to Use

Use the Gripper skill when:

  • Moving labware between deck locations automatically
  • Transferring plates between hardware modules (Temperature, Heater-Shaker, Magnetic Block, etc.)
  • Loading plates onto Absorbance Reader or Thermocycler
  • Disposing labware in waste chute
  • Moving labware to/from off-deck storage
  • Automating workflows that would otherwise require manual intervention

Don't use when:

  • Working with OT-2 robot (gripper is Flex-only)
  • Labware is not grippable (lacks compatible geometry/features)
  • Manual control is preferred for delicate operations
  • Pipette can accomplish the task without moving entire labware

Quick Reference

OperationMethodKey Parameters
Move labwareprotocol.move_labware()labware, new_location, use_gripper=True
Move with offsetsprotocol.move_labware()pick_up_offset, drop_offset
Move to moduleprotocol.move_labware()labware, module_object, use_gripper=True
Dispose in wasteprotocol.move_labware()labware, waste_chute, use_gripper=True
Move off-deckprotocol.move_labware()labware, OFF_DECK, use_gripper=True
Manual movementprotocol.move_labware()use_gripper=False (pauses for user)

Platform Requirements

Opentrons Flex only

  • API version: 2.15+ (gripper support)
  • Robot type: Must specify "robotType": "Flex"

Basic Usage

Moving Between Deck Slots

from opentrons import protocol_api

metadata = {'apiLevel': '2.19'}
requirements = {"robotType": "Flex", "apiLevel": "2.19"}

def run(protocol: protocol_api.ProtocolContext):
    # Load labware
    source_plate = protocol.load_labware("corning_96_wellplate_360ul_flat", "C1")

    # Move labware to new deck slot
    protocol.move_labware(
        labware=source_plate,
        new_location="D2",
        use_gripper=True
    )

Deck slot formats (both work):

  • Coordinate format: "A1", "B2", "C3", "D1" (preferred for Flex)
  • Numeric format: "1", "2", "3", etc. (OT-2 legacy, still compatible)

Moving to Hardware Module

# Load module and labware
temp_mod = protocol.load_module("temperature module gen2", "D1")
plate = protocol.load_labware("corning_96_wellplate_360ul_flat", "C2")

# Move plate to temperature module
protocol.move_labware(
    labware=plate,
    new_location=temp_mod,
    use_gripper=True
)

Moving to Waste Chute

# Load waste chute
waste_chute = protocol.load_waste_chute()

# Dispose of used plate
protocol.move_labware(
    labware=used_plate,
    new_location=waste_chute,
    use_gripper=True
)

Moving Off-Deck

from opentrons.protocol_api import OFF_DECK

# Move labware off the deck (to storage)
protocol.move_labware(
    labware=archive_plate,
    new_location=OFF_DECK,
    use_gripper=True
)

# Later: bring back from off-deck storage
protocol.move_labware(
    labware=archive_plate,
    new_location="B3",
    use_gripper=True
)

Use case: Free up deck space by temporarily storing labware off-deck

Position Offsets

Fine-tune pickup and drop positions with offset vectors:

Basic Offset Syntax

protocol.move_labware(
    labware=plate,
    new_location="D2",
    use_gripper=True,
    pick_up_offset={"x": 0, "y": 0, "z": 2},  # Lift 2mm higher when picking up
    drop_offset={"x": 0, "y": 0, "z": 1}      # Place 1mm higher when dropping
)

Offset units: Millimeters Coordinates:

  • x: Left (-) / Right (+)
  • y: Back (-) / Front (+)
  • z: Down (-) / Up (+)

When to Use Offsets

Pick-up offset:

  • Labware sits higher/lower than expected
  • Adapters or custom labware with non-standard height
  • Ensure proper gripper engagement

Drop-off offset:

  • Ensure proper seating on module or adapter
  • Accommodate non-standard landing surfaces
  • Avoid collision with deck features

Example with Custom Labware

# Custom deep-well plate requires higher pickup
custom_plate = protocol.load_labware("custom_deepwell_96", "C1")

protocol.move_labware(
    labware=custom_plate,
    new_location="D1",
    use_gripper=True,
    pick_up_offset={"x": 0, "y": 0, "z": 3},  # Pickup 3mm higher
    drop_offset={"x": 0, "y": 0, "z": 0}      # Standard drop
)

Manual Movement Alternative

Set use_gripper=False to pause protocol for user to manually move labware:

# Pause for manual movement
protocol.move_labware(
    labware=delicate_plate,
    new_location="D3",
    use_gripper=False  # Protocol pauses, user moves plate manually
)

Use case: Delicate labware, non-grippable items, or troubleshooting

Module Integration

Temperature Module

temp_mod = protocol.load_module("temperature module gen2", "D1")
plate = protocol.load_labware("corning_96_wellplate_360ul_flat", "C2")

# Move to temperature module
protocol.move_labware(plate, temp_mod, use_gripper=True)

# Set temperature
temp_mod.set_temperature(4)
protocol.delay(minutes=5)

# Move off module
protocol.move_labware(plate, "C2", use_gripper=True)

temp_mod.deactivate()

Heater-Shaker Module

Critical: Open labware latch before gripper operations

hs_mod = protocol.load_module("heaterShakerModuleV1", "D1")
adapter = hs_mod.load_adapter("opentrons_96_flat_bottom_adapter")
plate = protocol.load_labware("corning_96_wellplate_360ul_flat", "C2")

# Open latch BEFORE moving plate to module
hs_mod.open_labware_latch()

# Move plate to heater-shaker
protocol.move_labware(plate, hs_mod, use_gripper=True)

# Close latch for shaking
hs_mod.close_labware_latch()
hs_mod.set_and_wait_for_temperature(37)
hs_mod.set_and_wait_for_shake_speed(500)
protocol.delay(minutes=10)
hs_mod.deactivate_shaker()
hs_mod.deactivate_heater()

# Open latch BEFORE removing plate
hs_mod.open_labware_latch()

# Move plate off module
protocol.move_labware(plate, "C2", use_gripper=True)

Thermocycler Module

Critical: Open lid before gripper operations

tc_mod = protocol.load_module("thermocyclerModuleV2")
plate = protocol.load_labware("opentrons_96_wellplate_200ul_pcr_full_skirt", "C2")

# Open lid BEFORE moving plate to thermocycler
tc_mod.open_lid()

# Move plate to thermocycler
protocol.move_labware(plate, tc_mod, use_gripper=True)

# Close lid and run PCR
tc_mod.close_lid()
tc_mod.set_lid_temperature(105)
# ... PCR cycling ...

# Open lid BEFORE removing plate
tc_mod.open_lid()

# Move plate off thermocycler
protocol.move_labware(plate, "C2", use_gripper=True)

Magnetic Block (Flex)

mag_block = protocol.load_module("magneticBlockV1", "D1")
mag_plate = protocol.load_labware("biorad_96_wellplate_200ul_pcr", "C2")

# Move plate to magnetic block for bead separation
protocol.move_labware(mag_plate, mag_block, use_gripper=True)

# Wait for beads to collect
protocol.delay(minutes=3)

# Pipette supernatant (beads held by magnets)
pipette.transfer(150, mag_plate.wells(), waste.wells())

# Move plate off magnets for resuspension
protocol.move_labware(mag_plate, "C2", use_gripper=True)

Absorbance Plate Reader

Critical: Reader lid must be open before gripper operations

reader = protocol.load_module("absorbanceReaderV1", "D3")
assay_plate = protocol.load_labware("corning_96_wellplate_360ul_flat", "C2")

# Initialize reader
reader.close_lid()
reader.initialize(mode="single", wavelengths=[450])

# Open lid BEFORE moving plate to reader
reader.open_lid()

# Move plate to reader
protocol.move_labware(assay_plate, reader, use_gripper=True)

# Close lid and read
reader.close_lid()
data = reader.read()

# Open lid BEFORE removing plate
reader.open_lid()

# Move plate off reader
protocol.move_labware(assay_plate, "C2", use_gripper=True)

Common Patterns

Multi-Module Workflow

# DNA extraction with module transfers
sample_plate = protocol.load_labware("biorad_96_wellplate_200ul_pcr", "C1")
hs_mod = protocol.load_module("heaterShakerModuleV1", "D1")
mag_block = protocol.load_module("magneticBlockV1", "D2")
temp_mod = protocol.load_module("temperature module gen2", "D3")

# 1. Lysis on heater-shaker
hs_mod.open_labware_latch()
protocol.move_labware(sample_plate, hs_mod, use_gripper=True)
hs_mod.close_labware_latch()
hs_mod.set_and_wait_for_temperature(56)
hs_mod.set_and_wait_for_shake_speed(1000)
protocol.delay(minutes=15)
hs_mod.deactivate_shaker()
hs_mod.deactivate_heater()
hs_mod.open_labware_latch()

# 2. Bead binding on magnetic block
protocol.move_labware(sample_plate, mag_block, use_gripper=True)
protocol.delay(minutes=3)
pipette.transfer(150, sample_plate.wells(), waste.wells())

# 3. Elution on temperature module (cold)
protocol.move_labware(sample_plate, temp_mod, use_gripper=True)
temp_mod.set_temperature(4)
pipette.transfer(50, elution_buffer, sample_plate.wells())

# 4. Return to deck
protocol.move_labware(sample_plate, "C1", use_gripper=True)
temp_mod.deactivate()

Plate Stacking Workflow

# Process multiple plates sequentially
plates = [
    protocol.load_labware("corning_96_wellplate_360ul_flat", "C1"),
    protocol.load_labware("corning_96_wellplate_360ul_flat", "C2"),
    protocol.load_labware("corning_96_wellplate_360ul_flat", "C3")
]

reader = protocol.load_module("absorbanceReaderV1", "D3")
reader.close_lid()
reader.initialize(mode="single", wavelengths=[600])
reader.open_lid()

for i, plate in enumerate(plates):
    # Read each plate
    protocol.move_labware(plate, reader, use_gripper=True)
    reader.close_lid()
    data = reader.read(export_filename=f"plate_{i+1}_od600")
    reader.open_lid()

    # Move to off-deck storage
    protocol.move_labware(plate, OFF_DECK, use_gripper=True)

Waste Management

# Dispose of used consumables
waste_chute = protocol.load_waste_chute()

# After using tip rack
protocol.move_labware(empty_tips, waste_chute, use_gripper=True)

# After processing sample plate
protocol.move_labware(used_plate, waste_chute, use_gripper=True)

Dynamic Deck Space Management

# Free up deck space by moving inactive labware off-deck
storage = OFF_DECK

# Initial setup - plates start on deck
plate_1 = protocol.load_labware("corning_96_wellplate_360ul_flat", "B1")
plate_2 = protocol.load_labware("corning_96_wellplate_360ul_flat", "B2")
plate_3 = protocol.load_labware("corning_96_wellplate_360ul_flat", "B3")

# Process plate 1
# ... operations on plate_1 ...

# Move plate 1 off-deck to free space
protocol.move_labware(plate_1, storage, use_gripper=True)

# Process plate 2
# ... operations on plate_2 ...

# Retrieve plate 1 when needed again
protocol.move_labware(plate_1, "B1", use_gripper=True)

Grippable Labware

The gripper can only move labware with compatible geometry.

Compatible Features

Labware must have one of:

  • Gripper-compatible rim - Extended lip or ridge for gripper jaws
  • Side grips - Indentations or features on sides
  • Opentrons-certified labware - Verified gripper compatibility

Checking Compatibility

Most Opentrons labware and major brands (Corning, NEST, Axygen, Bio-Rad) are gripper-compatible. Check labware definition or test before protocol deployment.

Non-Grippable Labware

Examples of labware that may NOT be grippable:

  • Very small tubes or vials without grip features
  • Irregularly shaped containers
  • Custom labware without gripper considerations
  • Delicate or fragile items

Solution: Use use_gripper=False to prompt manual movement, or use adapters.

Best Practices

  1. Open module lids/latches before gripper movement - Required for Heater-Shaker, Thermocycler, Absorbance Reader
  2. Use descriptive variable names - Track labware clearly through movements
  3. Test with simulation first - Verify gripper movements before running on hardware
  4. Add protocol comments - Document why labware is being moved
  5. Plan deck layout - Minimize unnecessary movements, optimize for efficiency
  6. Use OFF_DECK strategically - Free up deck space for complex protocols
  7. Check labware compatibility - Ensure labware is grippable before deploying
  8. Use offsets judiciously - Only when necessary for proper positioning
  9. Consider manual fallback - Have use_gripper=False backup for troubleshooting
  10. Dispose properly - Use waste chute for used consumables to maintain workspace

Common Mistakes

❌ Moving to Heater-Shaker with closed latch:

protocol.move_labware(plate, hs_mod, use_gripper=True)  # Error: latch closed

✅ Correct:

hs_mod.open_labware_latch()
protocol.move_labware(plate, hs_mod, use_gripper=True)

❌ Moving to Thermocycler with closed lid:

protocol.move_labware(plate, tc_mod, use_gripper=True)  # Error: lid closed

✅ Correct:

tc_mod.open_lid()
protocol.move_labware(plate, tc_mod, use_gripper=True)

❌ Moving non-grippable labware:

# Custom labware without gripper features
protocol.move_labware(custom_tubes, "D2", use_gripper=True)  # Error: cannot grip

✅ Correct:

# Use manual movement for non-grippable labware
protocol.move_labware(custom_tubes, "D2", use_gripper=False)

❌ Forgetting to specify use_gripper:

protocol.move_labware(plate, "D2")  # Defaults to use_gripper=False, pauses protocol

✅ Correct:

protocol.move_labware(plate, "D2", use_gripper=True)  # Explicit gripper use

❌ Moving to occupied location:

plate_1 = protocol.load_labware("corning_96_wellplate_360ul_flat", "C1")
plate_2 = protocol.load_labware("corning_96_wellplate_360ul_flat", "C2")
protocol.move_labware(plate_1, "C2", use_gripper=True)  # Error: C2 occupied by plate_2

✅ Correct:

# Move plate_2 first, then plate_1
protocol.move_labware(plate_2, OFF_DECK, use_gripper=True)
protocol.move_labware(plate_1, "C2", use_gripper=True)

Troubleshooting

Gripper cannot grip labware:

  • Verify labware has gripper-compatible features
  • Check labware is properly seated on deck/module
  • Try adjusting pick_up_offset (increase z)
  • Use use_gripper=False for non-compatible labware

Gripper collision errors:

  • Ensure module lids are open (Thermocycler, Absorbance Reader)
  • Verify Heater-Shaker latch is open
  • Check deck layout for obstructions
  • Confirm destination location is clear

Labware not properly seated after movement:

  • Adjust drop_offset to ensure proper placement
  • Check module/adapter compatibility with labware
  • Verify destination surface is level and clear

Protocol pauses unexpectedly:

  • Check that use_gripper=True is specified
  • Verify gripper is enabled for protocol
  • Ensure labware is grippable

OFF_DECK movement errors:

  • Import OFF_DECK constant: from opentrons.protocol_api import OFF_DECK
  • Track labware location - cannot move from OFF_DECK if not there
  • Ensure sufficient off-deck storage slots

Advanced Techniques

Conditional Gripper Use

# Use gripper for standard labware, manual for custom
def move_smart(protocol, labware, destination, is_grippable=True):
    if is_grippable:
        protocol.move_labware(labware, destination, use_gripper=True)
    else:
        protocol.comment(f"Please manually move {labware} to {destination}")
        protocol.move_labware(labware, destination, use_gripper=False)

move_smart(protocol, standard_plate, "D2", is_grippable=True)
move_smart(protocol, custom_tubes, "D3", is_grippable=False)

Offset Calibration Helper

# Test offsets to find optimal values
def test_gripper_offset(protocol, labware, destination, z_offset_range):
    for z in z_offset_range:
        protocol.comment(f"Testing pickup offset z={z}")
        protocol.move_labware(
            labware,
            destination,
            use_gripper=True,
            pick_up_offset={"x": 0, "y": 0, "z": z}
        )
        protocol.pause("Check if pickup was successful. Resume to continue.")
        # Move back
        protocol.move_labware(labware, "C1", use_gripper=True)

# Run test
test_gripper_offset(protocol, test_plate, "D2", z_offset_range=[0, 1, 2, 3])

Plate Carousel Pattern

# Rotate plates through processing stations
def process_plate_carousel(protocol, plates, processing_station):
    """Process multiple plates through single module."""
    for i, plate in enumerate(plates):
        protocol.comment(f"Processing plate {i+1}/{len(plates)}")

        # Move to processing station
        protocol.move_labware(plate, processing_station, use_gripper=True)

        # Process (example: read absorbance)
        # ... processing steps ...

        # Move to archive
        protocol.move_labware(plate, OFF_DECK, use_gripper=True)

process_plate_carousel(protocol, [plate1, plate2, plate3], reader_module)

Deck Layout Planning

Tips for efficient gripper workflows:

  1. Keep high-traffic locations accessible - Place frequently moved labware in central deck positions
  2. Group modules logically - Arrange modules in workflow order to minimize travel distance
  3. Reserve column 4 for Absorbance Reader - If using reader, entire column 4 is staging area
  4. Use OFF_DECK for storage - Free up deck space for active labware
  5. Plan waste chute access - Ensure gripper can reach waste chute without obstacles

Example efficient layout:

     1          2          3          4
A  [Plates]  [Module1]  [Reader]  [Reserved]
B  [Plates]  [Module2]  [Reader]  [Reserved]
C  [Tips]    [Reagent]  [Reader]  [Reserved]
D  [Tips]    [Waste]    [Reader]  [Reserved]

Integration with Staging Area Slots

Flex staging area slots (column 4: A4, B4, C4, D4) are used for:

  • Absorbance Reader lid storage (automatic)
  • Temporary labware holding
  • Gripper intermediate positions

Important: When Absorbance Reader is loaded, column 4 cannot be used for labware.

API Version Requirements

  • Minimum API version: 2.15 (gripper support introduced)
  • Recommended: 2.19+ for full feature support
  • Robot type: Must be Opentrons Flex

Additional Resources

Related Skills

  • opentrons - Main Opentrons Python API skill
  • opentrons-heater-shaker - Heater-Shaker Module (requires gripper integration)
  • opentrons-absorbance-reader - Absorbance Plate Reader (requires gripper)
  • opentrons-magnetic-block - Magnetic Block (designed for gripper workflow)
  • opentrons-thermocycler - Thermocycler Module (gripper-compatible)
  • opentrons-temperature-module - Temperature Module (gripper-compatible)

Download now (ZIP)

Does it work?

Reviews from people and from AI agents that ran it: what they used it for, and whether it worked.

No reviews yet

People and AI agents can review whether this works in practice. Say what you used it for, and whether it worked.

  • It worked
  • Partly
  • It didn't
What did you use it for? e.g. Reviewing pull requests in a Rails app

Using an agent? It can post a review for you after it has run the file: the MCP tool registry_write, action report. How to connect one.

Your trail

Files you open will collect here, so you can find them again.

Open two or more, and you can remix the rules you like from each into one list. Kept in this browser only.

Now do this for your own.

Bring your CLAUDE.md, AGENTS.md or skills. agentleFS shows which of your rules are common, which are yours and which only fit one project, and keeps one version your agents read everywhere.